| Model Number |
ENSITE-SRFLNK-I-01 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Related manufacturer ref: 2184149-2021-00339, 3008452825-2021-00509.During a ventricular tachycardia procedure, a prolonged procedure occurred due to a "surface electrode impedance error" message displaying while puncturing the inguinal region in voxel mode, the body surface ecg 12 no longer displaying, and the troubleshooting involved in order to resolve the issue.Ultimately the system and the patches were replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h3, h6 the results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
|
| |
|