• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE¿ X EP SYSTEM SURFACELINK MODULE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE¿ X EP SYSTEM SURFACELINK MODULE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-SRFLNK-I-01
Medical Device Problem Codes Communication or Transmission Problem (2896); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturer ref: 2184149-2021-00339, 3008452825-2021-00509.During a ventricular tachycardia procedure, a prolonged procedure occurred due to a "surface electrode impedance error" message displaying while puncturing the inguinal region in voxel mode, the body surface ecg 12 no longer displaying, and the troubleshooting involved in order to resolve the issue.Ultimately the system and the patches were replaced and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Additional information: g3, h2, h3, h6 the results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ X EP SYSTEM SURFACELINK MODULE
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12770262
Report Number2184149-2021-00340
Device Sequence Number3343109
Product Code DQK
UDI-Device Identifier05415067033611
UDI-Public(01)05415067033611(10)788110001
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202066
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberENSITE-SRFLNK-I-01
Device Catalogue NumberENSITE-SRFLNK-01
Device Lot Number788110001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2021
Supplement Date Received by Manufacturer11/29/2021
Initial Report FDA Received Date11/08/2021
Supplement Report FDA Received Date11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/11/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENSITE X EP SYSTEM SURFACE ELECTRODE KIT ; ENSITE X  EP SYSTEM AMPLIFIER
Patient Age83 YR
Patient SexMale
-
-