BIOMET MICROFIXATION 2.7X8MM HT X-DRIVE SCREW; CRANIOFACIAL BONE SCREW, NON-BIOABSORBABLE, STERILE Back to Search Results
Catalog Number 91-2708
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the fossa prosthesis and mandibular component never fixed to the desired trajectory upon implantation.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: tmj med rtfossa comp cat# 24-6560 lot# 7951900 45mm rt standard mandibular cat# 24-6545 lot# 829250c 2.7x8mm ht x-drive screw cat# 91-2708 lot# 325450 2.7x10mm ht x-drive screw cat# 91-2710 lot# 398050.(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2021-00517, 0001032347-2021-00518, 0001032347-2021-00519, 0001032347-2021-00520.
 
Additional Manufacturer Narrative
The 2.7x8mm ht x-drive screw (item# 91-2708, lot# 325450) was returned for investigation.The screw appears to be bent/warped.This was checked by rolling it across a flat table.The screw showed signs of use, with damaged threads and wear on the head.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.It is alleged that the devices were implanted incorrectly, however, without medical records a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7X8MM HT X-DRIVE SCREW
Common Device Name
CRANIOFACIAL BONE SCREW, NON-BIOABSORBABLE, STERILE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer delaney
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key12770286
Report Number0001032347-2021-00521
Device Sequence Number14534511
Product Code HWC
UDI-Device Identifier00841036053635
UDI-Public(01)00841036053635(10)325450
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K001238
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number91-2708
Device Lot Number325450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2021
Supplement Date Received by Manufacturer01/21/2022
03/21/2022
Initial Report FDA Received Date11/08/2021
Supplement Report FDA Received Date02/14/2022
03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-