• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX AIRWORTHY; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX AIRWORTHY; DEFIBRILLATOR Back to Search Results
Model Number M3536M
Medical Device Problem Code Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to philips that the device experienced a co2 calibration failure.The customer requested assistance from a philips remote service engineer (rse).The customer explained that their unit was not showing a co2 value and subsequently failing operational testing.Based on the noted issue, the rse advised the customer to swap out the co2 cable in an attempt to troubleshoot the issue.Should the issue remain, the customer was then instructed to replace the etco2 module.Based on the conclusion of the investigation, it was determined that this was a malfunction of the co2 cable.The customer followed the instruction provided by the rse and verified that swapping out the co2 cable resolved the reported failure.The device remains at the customer site.There is no indication of a systemic problem.No further investigation or action is warranted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX AIRWORTHY
Common Device Name
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
jessica ruuth
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12771830
Report Number3030677-2021-15443
Device Sequence Number4254260
Product Code MKJ
UDI-Device Identifier00884838006676
UDI-Public00884838006676
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K031187
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM3536M
Device Catalogue NumberM3536M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2021
Initial Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-