• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problems Overheating of Device (1437); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during set up, the mdu got hot and then froze.The procedure was successfully completed without delay using a s+n back-up device.No patient complications were reported.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A visual inspection found no issues.A functional evaluation revealed a handpiece sensor fault.Overheating of the device wasn't confirmed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat event.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint of overheating was not confirmed.Factors that could have contributed to the reported event include a failure of a concomitant device.The complaint of a handpiece sensor fault was confirmed, and the root cause has been associated with an electrical component failure.Factors that could have contributed to the reported event include a failure of the reed switch or damage on the hall board which may contribute to a short circuit.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12774663
MDR Text Key282713839
Report Number1643264-2021-02400
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2021
Initial Date FDA Received11/08/2021
Supplement Dates Manufacturer Received03/31/2022
Supplement Dates FDA Received04/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-