MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING SL; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24702
Medical Device Problem Codes Material Puncture/Hole (1504); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the guidewire perforated the catheter shaft.A 2.0mm x 100mm x 150cm sterling sl balloon catheter was selected for percutaneous transluminal angioplasty (pta).The 99% stenosed, severely calcified and severely tortuous lesion was located in an artery below the knee (btk).There was difficulty advancing the balloon through the lesion.The balloon was removed to exchange for a lower profile balloon.During removal, they observed the non-boston scientific guidewire was protruding through the shaft proximal to the balloon.The procedure was completed with another device.There were no patient complications.
 
Additional Manufacturer Narrative
Device eval by mfr: the returned product consisted of a sterling sl balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the shaft.Microscopic examination revealed a hole in the guidewire lumen and inflation lumen 34mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that the guidewire perforated the catheter shaft.A 2.0mm x 100mm x 150cm sterling sl balloon catheter was selected for percutaneous transluminal angioplasty (pta).The 99% stenosed, severely calcified and severely tortuous lesion was located in an artery below the knee (btk).There was difficulty advancing the balloon through the lesion.The balloon was removed to exchange for a lower profile balloon.During removal, they observed the non-boston scientific guidewire was protruding through the shaft proximal to the balloon.The procedure was completed with another device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING SL
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12774873
Report Number2134265-2021-14009
Device Sequence Number5657724
Product Code DQY
UDI-Device Identifier08714729782216
UDI-Public08714729782216
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K093720
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2023
Device Model Number24702
Device Catalogue Number24702
Device Lot Number0026710062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2021
Supplement Date Received by Manufacturer12/08/2021
Initial Report FDA Received Date11/08/2021
Supplement Report FDA Received Date12/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-