MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3560
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the burr became stuck on the rotawire.A 1.50mm rotalink and a rotawire were selected to be used for an angioplasty procedure.The rotapro was prepped and platformed at 160000rpm outside the patient,then advanced over the wire.During procedure, there was significant resistance between the burr and rotawire during advancement so that at a point the physician was unable to advance the burr on the wire towards the lesion.Then, the rotawire got stuck in the rotalink burr.The rotapro and the rotawire were removed together as one unit from the patient.The procedure completed successfully with another 1.50 mm burr and rotawire devices.There were no patient complications and the patient was good condition post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE AND WIRECLIP TORQUER
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12774958
Report Number2134265-2021-14004
Device Sequence Number17037138
Product Code MCX
UDI-Device Identifier08714729144168
UDI-Public08714729144168
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2022
Device Model Number3560
Device Catalogue Number3560
Device Lot Number0026537263
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2021
Initial Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-