• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN TRUSEAL ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN TRUSEAL ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635TU70040
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sheath kink occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman truseal access system (was) was positioned and a watchman laa closure device & delivery system (wds) were used.During deployment of the watchman device, the was sheath was torqued and became kinked.Upon recapture of the device, the kink worsened.Additional intervention was required to recapture the watchman device and remove the kinked sheath.The back end of the deployment knob of the device was cut to remove the sheath and exchanged for a new was sheath.The watchman device was recaptured with the use of the new was by pulling the back end of the core wire which pulled the device into the new was sheath.Ultimately the patient was successfully implanted with a watchman device and reported to have been discharged the next day without issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN TRUSEAL ACCESS SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12775176
Report Number2134265-2021-13919
Device Sequence Number12268315
Product Code DQY
UDI-Device Identifier08714729965725
UDI-Public08714729965725
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/26/2024
Device Model NumberM635TU70040
Device Catalogue NumberM635TU70040
Device Lot Number0027918503
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/22/2021
Initial Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight62 KG
-
-