The investigation determined that a higher than expected ammonia (amon) result was obtained when the customer processed a non-vitros (mas) quality control (qc) fluid using vitros amon lot 1018-0256-9804 on a vitros xt7600 integrated system.A definitive assignable cause of the event was not determined.An instrument issue cannot be ruled out as a contributor to the event as no diagnostic precision testing was conducted around the time of the event ((b)(6) 2021); therefore it cannot be determined whether the instrument was performing as expected when the higher than expected result was obtained.Qc results following ortho field engineer service actions performed on 13 october 2021 were acceptable in terms of amon accuracy and precision.However, since the service was over 5 weeks after the day of the event, it cannot be confirmed or ruled out if an instrument issue contributed to the event.A vitros amon lot 1018-0256-9804 issue cannot be ruled out as a contributor to the event, as historical qc results indicated unacceptable precision leading up to the event on (b)(6) 2021.However, ongoing tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros amon lot 1018-0256-9804.(b)(4).
|