• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.8MM KIRSCHNER WIRE SPADE POINT 200MM; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.8MM KIRSCHNER WIRE SPADE POINT 200MM; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.108.005
Medical Device Problem Codes Material Deformation (2976); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/ investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, during a left lower limb reconstructive surgery, two (2) 2.8mm k-wires stripped and partially bent over their entire length and one (1) 2.0mm k-wire bent and stripped.No further information provided.This report is for one (1) 2.8mm kirschner wire spade point 200mm.This is report 2 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d10 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6 - codes updated to imdrf codes.Investigation summary: the complaint device was not received for investigation.A photo investigation was performed based on the image attached in the email in notes & attachments section of pc titled "source file - formato de novedades ktfm clínica noel.Colectivo 286417.Pdf".The image was reviewed, and the complaint condition is not confirmed.The image provided does not show any damage to the spade tip k-wire, therefore it cannot be confirmed that it is bent or scratched.A definitive assignable root cause could not be determined based on the provided information.As the device was not returned, an as-received condition could not be assessed and a dimensional inspection and document/specification review were not completed.During the investigation, no product design issues or discrepancies were observed (based on the image) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Background; procedure: left lower limb reconstructive surgery.Visual inspection: the k-wire ø2.8 w/spade point tip (p/n: 03.108.005 & lot #: 93p5680) was returned and received at us cq.Upon visual inspection, it was observed that k-wire is little bent and had scratches all across the surface.No other issues were identified with the returned device.Document/specification review based on the date of manufacture the following drawings, reflecting the current and manufactured revision were reviewed investigation conclusion the complaint condition was confirmed for the device.No definitive root cause can be assigned based on the available information.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device history lot a manufacturing record evaluation was performed for the finished device 03.108.005 lot number 93p5680, and no non-conformance's were identified.Device history review a manufacturing record evaluation was performed for the finished device 03.108.005 lot number 93p5680, and no non-conformance's were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM KIRSCHNER WIRE SPADE POINT 200MM
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BALSTHAL (CH)
dornacherstrasse 20
balsthal 4710
SZ   4710
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12775273
Report Number8030965-2021-09380
Device Sequence Number19734921
Product Code FZX
UDI-Device Identifier07611819847255
UDI-Public(01)07611819847255
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.108.005
Device Lot Number93P5680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2021
Supplement Date Received by Manufacturer11/17/2021
12/29/2021
Initial Report FDA Received Date11/08/2021
Supplement Report FDA Received Date11/17/2021
12/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
K-WIRE Ø2 L280 SST; K-WIRE Ø2.8 W/SPADE POINT TIP
Patient SexFemale
-
-