• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2021
Event Type  malfunction  
Manufacturer Narrative
Customer does not have any additional information to provide.The device is returned and an evaluation completed for it.Upon inspection and testing, it was observed that the bending section was broken with metal sticking out.The customer¿s complaint of broken bending section was confirmed.Additionally, it was noted that the distal end rubber insulation was leaking and the drum unit was corroded.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event Description
Customer returned the device for an evaluation of bending section breakage.There is no reported harm to any patient.Upon evaluation of the returned device, it was noted that the bending section skeleton was broken with metal sticking out.This medwatch is being submitted for the reportable malfunction of the broken bending section skeleton with metal protruding.
 
Manufacturer Narrative
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.And to correct information provided on the initial report.The following sections were corrected: g2, a review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it was surmised, that excessive force was applied through user manipulation of the bending section, resulting to damage of the bending tube.The instruction manual identifies, the following related verbiage: ¿precautions: perform a leakage test on the endoscope, after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope, may also pose an infection control risk.Inspection of the endoscope: inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities¿.The instruction manual identifies the following, additional related verbiage that may reduce/prevent occurrence of the suggested event by device handling: ¿do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction, while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12775296
MDR Text Key282723050
Report Number8010047-2021-14282
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-