| Model Number |
2800 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Regurgitation which develops progressively over time can be due to a number of issues including patient related factors or structural valve deterioration.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Regurgitation may also develop progressively if host fibrotic tissue, or pannus, grows onto the bioprosthetic valve.Pannus, a cause of nonstructural dysfunction, may interfere with functionality of the device by restricting the leaflet motion leading to abnormal coaptation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.The subject device is not available for evaluation as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 16 years, six (6) months due to insufficiency.The tavr was performed with a 26mm 9750tfx transcatheter valve with no complications.No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: updated: b5 h11: corrected data: based on the additional information obtained, this event is no longer considered reportable, and this correction is being submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 16 years, six (6) months due to insufficiency.The tavr was performed with a 26mm 9750tfx transcatheter valve with no complications.According to the physician, there was no allegation that the surgical valve prematurely failed.
|
| |
|