Model Number HST III SYSTEM (4.3MM) |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/08/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Trackwise id# (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
|
Event Description
|
The hospital reported that during a coronary artery bypass procedure using hst iii system (4.3mm), the blocking umbrella fell off during loading.Replace the same equipment to complete the operation - no effect on patients.
|
|
Manufacturer Narrative
|
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure mode.Historical data analysis: (4109/213/67) there was 3 additional similar complaint reported for the reported failure mode and the reported lot/serial number.A review of the product complaint trend data was performed for the months of ¿jul 2021¿ through ¿oct 2021¿.The review does not identify an increasing trend or an adverse trend for three consecutive months for the product family and the failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period nov 2019 through oct 2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
|
|
Search Alerts/Recalls
|