• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HST III SYSTEM (4.3MM); CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HST III SYSTEM (4.3MM); CLAMP, VASCULAR Back to Search Results
Model Number HST III SYSTEM (4.3MM)
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2020
Event Type  malfunction  
Manufacturer Narrative
Trackwise id# (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event Description
The hospital reported that during a coronary artery bypass procedure using hst iii system (4.3mm), the blocking umbrella fell off during loading.Replace the same equipment to complete the operation - no effect on patients.
 
Manufacturer Narrative
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure mode.Historical data analysis: (4109/213/67) there was 3 additional similar complaint reported for the reported failure mode and the reported lot/serial number.A review of the product complaint trend data was performed for the months of ¿jul 2021¿ through ¿oct 2021¿.The review does not identify an increasing trend or an adverse trend for three consecutive months for the product family and the failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period nov 2019 through oct 2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III SYSTEM (4.3MM)
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key12776160
MDR Text Key280818725
Report Number2242352-2021-00925
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700321
UDI-Public00607567700321
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/08/2021
Device Model NumberHST III SYSTEM (4.3MM)
Device Catalogue NumberHSK-3043
Device Lot Number25151996
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/12/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/05/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight68 KG
-
-