Model Number 3CX*RX15RE40
Medical Device Problem Code Clumping in Device or Device Ingredient (1095)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, there were clots found in the oxygenator and the blood gases were dropping rapidly.They were on by-pass for about 3 hours.They were about to come off bypass in about 5 minutes.That is when bioglue was administered to the patient.Within about 5 minutes of bioglue administration, little clots were noticed in the oxygenator and the blood gases were dropping rapidly.The customer feels strongly that the bioglue got into the circuit and caused the issues.There was approximately 240 ml of blood loss.The product was changed out.The surgery was completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 9, 2021.Upon further investigation of the reported event, the following information is new and/or changed: g6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H6 (identification of evaluation codes 10, 11, 3331, 213, 67).Type of investigation 1#: 10 - testing of actual/suspected device.Type of investigation 2#: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation 3#: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The actual sample was visually inspected and did not find any anomaly including breakage.After rinsing and drying, the sample was installed into a circuit consisting of tubes, and the pressure drop when circulating bovine blood (hct: 35%, temp: 37 degrees) was measured.It was confirmed to meet the factory's specifications and no obstruction was found.Based on the investigation result, no anomaly that could lead to clot formation was found in the actual sample.Review of the manufacturing record and the incoming inspection record of the actual sample confirmed that there was no anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot# combination.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLLOW FIBER OXY WITH 4000 M
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key12777974
Report Number1124841-2021-00255
Device Sequence Number5942619
Product Code DTZ
UDI-Device Identifier00699753450158
UDI-Public(01)00699753450158
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K153213
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number3CX*RX15RE40
Device Catalogue NumberN/A
Device Lot NumberZE12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2021
Supplement Date Received by Manufacturer01/03/2022
Initial Report FDA Received Date11/09/2021
Supplement Report FDA Received Date01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-