| Model Number |
MP5313-C |
| Medical Device Problem Codes |
Backflow (1064); Complete Blockage (1094); Free or Unrestricted Flow (2945)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd maxplus¿ pressure rated extension set with removable needleless connector check valve malfunction failed, causing flow issues and blood backflow during use.The following information was provided by the initial reporter: "this issue is with the pigtail iv set.The rubber inside the valve is not closing in the opening after use.The vacutainer is inserted, i withdraw blood but when i am done and try to disconnect the rubber inside is not recoiling and closing the valve, leaving blood to backflow.".
|
| |
|
Additional Manufacturer Narrative
|
|
Date of birth: unknown.The patient¿s age was used to determine a placeholder date for this field.Fda notified?: the initial reporter also notified the fda via medwatch # (b)(4).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: no product or photo was returned by the customer.It was reported that the valve is not closing leaving the blood to backflow.The customer complaint could not be verified due to the product not being returned for failure investigation.A device history record review for model mp5313-c lot number 20126036 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.A complaint history check was performed and this is the 1st related complaint reported with the defect/condition of flow issues - back flow with lot #20126036 regarding item #mp5313-c.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd maxplus¿ pressure rated extension set with removable needleless connector check valve malfunction failed, causing flow issues and blood backflow during use.The following information was provided by the initial reporter: "this issue is with the pigtail iv set.The rubber inside the valve is not closing in the opening after use.The vacutainer is inserted, i withdraw blood but when i am done and try to disconnect the rubber inside is not recoiling and closing the valve, leaving blood to backflow.".
|
| |
|