| Model Number |
38265 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: imager ii/5/straight/65/035 sgl was returned for analysis.Returned product to boston scientific consisted of an imager diagnostic catheter in an unopened sterile pouch.The device was examined for damage or any irregularities.The sterile pouch showed damage and a hole was observed.The sterile pouch was opened, and the device showed a smashed hub matching where the pouch damage was.No other damage was noticed.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was confirmed for hub damage as well as sterile pouch damage.
|
| |
|
Event or Problem Description
|
|
It was reported that hub damaged occurred.An imager ii angiographic catheter was selected for use.During preparation, upon opening the package they noticed the hub was smashed and they were unable to use it.The procedure was completed with an alternate method/device.There were no patient complications reported.However, returned device analysis revealed sterile pouch damage.
|
| |
|