MAUDE Adverse Event Report: ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problems Battery Problem (2885); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.During processing of this incident, attempts were made to obtain complete event information.Further information was requested but not received.
 
Event Description
It was reported that the patient's ipg was nearing end of life.In turn, surgical intervention was undertaken wherein the ipg was explanted and replaced, resolving the issue.
 
Event Description
Device analysis revealed that the ipg had reached normal eri.No allegations against the ipg.
 
Manufacturer Narrative
The report of ipg approaching end of life was confirmed.Analysis of the ipg found the device declared eri and had normal battery depletion.H6 - adverse event problem - corrected device code to reflect normal eri.There is no allegation against the ipg, the ipg is no longer reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12778490
MDR Text Key280516224
Report Number1627487-2021-18089
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2019
Device Model Number6662
Device Catalogue Number6662
Device Lot Number6115384
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-