MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ESCAPE; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ESCAPE; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063902010
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an escape basket was unpacked on (b)(6) 2021 for a ureteroscopy.During preparation, a hole was noticed in the packaging and the sterility of the device was compromised.Another escape basket was used to complete the procedure.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
This event was reported by the materials manager.(b)(6).(b)(4).The device has been received for analysis.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an escape basket was unpacked on (b)(6) 2021 for a ureteroscopy.During preparation, a hole was noticed in the packaging and the sterility of the device was compromised.Another escape basket was used to complete the procedure.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Block e1: this event was reported by the materials manager.The physician present for this case was: (b)(6).Device code a020504 captures the reportable event of hole in the packaging.Block h10: the returned escape basket was analyzed, and a visual evaluation noted that the pouch was torn in one of the sides.The unit did not present issues due to was inside of the original pouch and the tray looks closed.No other issues was noted.Based on all available information, the most probable root cause is cause not established since the investigation findings do not lead to a clear conclusion about the cause of the reported event.There is no evidence of a manufacturing issue, design or user issue which could have caused the complaint.As per external support form it was concluded that the manufacturing process has enough controls to assure the product performance.Most likely procedural factors as handling of the device boxes prior to use, could have led to the reported event.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A search of the complaint database confirmed that no similar complaints exist for the specified lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESCAPE
Common Device Name
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12779025
Report Number3005099803-2021-05800
Device Sequence Number14958759
Product Code FFL
UDI-Device Identifier08714729121299
UDI-Public08714729121299
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/23/2023
Device Model NumberM0063902010
Device Catalogue Number390-201
Device Lot Number0025759741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/18/2021
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date11/09/2021
Supplement Report FDA Received Date11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-