• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. FLEX SHOULDER SYSTEM REVERSED; PROSTHESIS SHOULDER METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. FLEX SHOULDER SYSTEM REVERSED; PROSTHESIS SHOULDER METAL Back to Search Results
Model Number TRAY DIA. 40 TH 0 ECC 1.5
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is related to voluntary recall fa- wmg-2021-005 for product mix.During reassessment of the reported event after health hazard evaluation assessment was concluded the reportability decision was revised to be reported under the vigilance process with the awareness date of october-14-2021.All affected parts were recalled.Investigation was conducted and confirmed that two specific lots of flex reversed tray have been swapped and contain the incorrect product in the package : aequalis ascend flex reversed tray (+)0 thickness, 1.5mm offset, reference dwf510, lot 1197aw and aequalis ascend flex reversed tray (+)0 thickness, 0.0mm offset, reference dwf500, lot 1179aw a review of the device history for the reported lot did not indicate any abnormalities.Based on investigation, the root cause was attributed to a manufacturing related issue.The affected parts were recalled.The root cause is attributed to a manufacturing defect (human error during a packaging step).
 
Event or Problem Description
The implants in both boxes do not correspond to the label on each package/box.The surgeon planned a ascend flex reversed tray low 1.5/+0mm implant and got by opening the first implant box an ascend flex centered tray instead.Due to this rejected pack a second implant (dwf510) was opened with inside another ascend flex centered tray implant identical to the one packed in the first implant box.Surgery was completed by using and implanting an ascend flex high 3.5/+0mm reversed tray.It delayed the surgery of max 10-15 min.No additional anesthesia was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEX SHOULDER SYSTEM REVERSED
Common Device Name
PROSTHESIS SHOULDER METAL
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint martin, 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint martin, 38330
FR   38330
Manufacturer Contact
laurent ray
161 rue lavoisier
montbonnot saint martin, 38330
FR   38330
MDR Report Key12779139
Report Number3000931034-2021-00050
Device Sequence Number3608194
Product Code KWS
UDI-Device Identifier03700386941054
UDI-Public03700386941054
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K122698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2026
Device Model NumberTRAY DIA. 40 TH 0 ECC 1.5
Device Catalogue NumberDWF510
Device Lot Number1197AW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2021
Initial Report FDA Received Date11/09/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FA- WMG-2021-005
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-