| Model Number |
2975-00-720 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the push rod tip was bent, forced into trq shft 95, and now stuck together.Locking tip missing from body collet.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will bereviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: the device associated with this report was returned for examination.The investigation found sufficient evidence to confirm, the reported event.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|