Model Number 74016424
Device Problems Detachment of Device or Device Component (2907); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, during a tka surgery, the pin on side of journey ii uni femoraltrial holder holding the spring loaded clamps to the size of the impactor fell out.The hook, spring and pin were all recovered from the patient with tweezers and fingers.Surgery was resumed, without any delay, by changing the surgical technique.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned jrny ii uni femoraltrial holder confirms one of the femoral trail holder arms broke off.The broken piece was returned with the device.The device exhibits signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II UNI FEMORALTRIAL HOLDER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12780026
MDR Text Key280530098
Report Number1020279-2021-07927
Device Sequence Number1
Product Code LXH
UDI-Device Identifier00885556692264
UDI-Public00885556692264
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74016424
Device Catalogue Number74016424
Device Lot Number18KSC0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-