|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted some of the staples were partially advanced in the cartridge.The staple pushers were reset and the staples were applied with proper staple formation.Functionally, since the instrument was not received a representative instrument 0409 was used for testing.The usable advanced staples in the sulu were reset inside the cartridge.The instrument and the single use loading unit (sulu) were applied to appropriate test media.The jaw closes smoothly, the pin slide advances fully, and the handle actuates smoothly into pre-clamped and clamped positions.The jaw opens, handle unlocks and pin slide retracts when black release button is depressed.Acceptable staple formation was observed on test media.It was reported that the device did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue occurs when the loading unit is inadvertently pushed up after the firing cycle is started or improperly loaded prior to application.In these cases, only the bottom set of pushers are in contact with the thrust bar, leaving the top set of pushers unfired.The manufacturing records for each device are thoroug hly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to reload the instrument, grasp the new, unused loading unit by the finger pads, with the staple holes facing the instrument anvil.Insert the loading unit into the metal cartridge housing and push firmly downward until the single use loading unit clicks into position.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|