Additional information: d9, g2, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the loading unit had been partially applied.Functionally, a representative instrument was used for testing.The advanced staples in the single use loading unit were reused and reset inside the cartridge.The representative instrument and the single use loading unit were applied to appropriate test media.The jaw closed smoothly, the pin slide advanced fully, and the handle actuated smoothly into pre-clamped and clamped positions.The jaw opened, handle unlocked, and pin slide retracted when black release button was depressed.Acceptable staple formation was observed on test media.It was reported that the device did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the ins trument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|