| Model Number |
1120912 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Respiratory Arrest (4461)
|
| Date of Event |
02/05/2021
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The manufacturer was made aware of a patient that was set up on a trilogy evo ventilator and an af 541 mask for covid pneumonia and copd.It was alleged that the patient was discovered in an unresponsive condition, with the tube connecting her face mask to a bipap ventilator detached at the connection point to the mask.It was reported the patient later expired.The manufacturer is currently investigating this event and a follow up report will be filed when the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The af541 mask in use was not retained due to covid-19 contamination.The af541 mask elbow design is a 22mm non-latching conical connection compliant with iso 5356-1.This design is not intended to prevent accidental disconnection, and the af541 mask is contraindicated for use with patients requiring life support ventilation.Based on the information provided, respironics concludes that the af541 mask operated as designed.
|
| |
|