• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EXTENSION, DUAL 4 CHANNEL 30CM; SCS EXTENSION Back to Search Results
Model Number 3343
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 10/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Related manufacturer report numbers 1627487-2021-18229, 1627487-2021-18230.It was reported that the patient was not receiving effective therapy from their system.It was found that the extensions had high impedances.In turn, surgical intervention was undertaken wherein the extensions were explanted and replaced.Postoperatively, the patient has effective therapy.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 30CM
Common Device Name
SCS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12780423
Report Number1627487-2021-18231
Device Sequence Number3608234
Product Code LGW
UDI-Device Identifier05414734402293
UDI-Public05414734402293
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2013
Device Model Number3343
Device Catalogue Number3343
Device Lot Number3346026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2021
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date11/09/2021
Supplement Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3341, SCS EXTENSION; 3343, SCS EXTENSION
Outcome Attributed to Adverse Event Other;
Patient Age31 YR
Patient SexFemale
Patient Weight59 KG
-
-