• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Model Number 66024008
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reference number: (b)(4).
 
Event or Problem Description
It was reported that , during treatment, a dressing of iv3000 1 hand 10x12cm ctn 50 was creased and detached with the backing paper.Treatment was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: we have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issues highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.The complaint history file contains a small number of further instances of the reported event in the last 3 years.It was reported that the dressings were used for treatment and creasing was found on the film.Samples returned for evaluation were examined and found to be creased and lifted from the final back tab paper as described in the complaint.This confirmed a relationship between the reported event and the device.There are checks in place to prevent creasing of the film.If creasing is identified during in-process checks, it is tabbed and recorded in the fault log.Minor creasing can sometimes occur in the film during the adhesive application process at the start or end of the roll.As in-process checks failed to identify creasing to the film on this occasion, the most probable root cause was determined as a manufacturing process issue.The alleged failure mode and any associated harm is mitigated in the risk files for this product this investigation is now complete with no corrective actions required.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 10X12CM CTN 50
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12780435
Report Number8043484-2021-01941
Device Sequence Number19759950
Product Code KGX
UDI-Device Identifier05000223417406
UDI-Public05000223417406
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2023
Device Model Number66024008
Device Catalogue Number4008
Device Lot Number2032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2021
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date11/09/2021
Supplement Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-