| Model Number |
L20H-002X02 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The excor driving tube (red) with lot number 00094146 was used from (b)(6) 2021 to (b)(6) 2021 (39 days).We have reviewed the production records of the excor driving tube with the lot number 00094146.This driving tube was produced according to our specification.A detailed investigation report will be provided as soon as it is available.
|
| |
|
Event or Problem Description
|
|
Berlin heart (b)(4) was informed by the clinic that a leakage was detected in the driving tube of the excor blood pump of a patient supported in the lvad configuration.The driving tube was exchanged in the clinic.The exchange was performed without complications.The patient was not negatively affected by the exchange and the patient is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
The excor driving tube (red) with lot number 00094146 was used from (b)(6) 2021 to (b)(6) 2021 (39 days).We have reviewed the production records of the excor driving tube with the lot number 00094146.This driving tube was produced according to our specification.The clinic has informed us that the driving tube in question was already disposed.Therefore berlin heart could not analyse the product in question.The clinic did not provide us with further information such as pictures or videos of the defect.Due to the fact, that the product was not available for analysis we can not drew any conclusion regarding the complained defect nor its cause.
|
| |
|