ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012447-06
Medical Device Problem Codes Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that following prepping and flushing the device, the yellow protective sheath of the 2.5x6.0mm nc trek balloon dilatation catheter (bdc) could not be removed.The device was not used and there was no patient involvement.A 2.50x15mm nc trek bdc was used to complete the procedure.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.(b)(4).
 
Additional Manufacturer Narrative
Visual, dimensional and functional inspections were performed on the returned device.The reported difficulty removing the protective sheath was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints from this lot.It was reported that the bdc was prepped prior to the sheath being removed.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instructions for use (ifu) states: complete the following steps to prepare the nc trek rx coronary dilatation catheter for use: slide the protective sheath off the balloon prior to performing air aspiration.In this case, it is unknown if the violation of the ifu caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported difficulty removing the protective sheath.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12781290
Report Number2024168-2021-10177
Device Sequence Number14958773
Product Code LOX
UDI-Device Identifier08717648151736
UDI-Public08717648151736
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number1012447-06
Device Catalogue Number1012447-06
Device Lot Number90628G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/15/2021
Supplement Date Received by Manufacturer12/08/2021
Initial Report FDA Received Date11/09/2021
Supplement Report FDA Received Date12/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-