| Model Number |
1012447-06 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that following prepping and flushing the device, the yellow protective sheath of the 2.5x6.0mm nc trek balloon dilatation catheter (bdc) could not be removed.The device was not used and there was no patient involvement.A 2.50x15mm nc trek bdc was used to complete the procedure.There was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional and functional inspections were performed on the returned device.The reported difficulty removing the protective sheath was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints from this lot.It was reported that the bdc was prepped prior to the sheath being removed.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instructions for use (ifu) states: complete the following steps to prepare the nc trek rx coronary dilatation catheter for use: slide the protective sheath off the balloon prior to performing air aspiration.In this case, it is unknown if the violation of the ifu caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported difficulty removing the protective sheath.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|