MEDTRONIC MEXICO KYPHON XPANDER¿ INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Model Number KPT1502
Device Problem Material Rupture (1546)
Patient Problem Rupture (2208)
Event Date 10/15/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with spinal therapy.It was reported that the instrument was damaged when inflated.The rupture occurred during inflation.There was a delay of less than 60 min in overall procedure time.The procedure was performed successfully by using a new product.There was no fragment left in the patient.There were no symptoms reported.There were no further complications reported regarding the event.
 
Manufacturer Narrative
H3: product analysis: part# kpt1502, lot# 222517223 visual and functional inspection revealed the balloon has been damaged.The damage is a slice/puncture in the middle portion of the balloon.This type of damage is consistent with the balloon coming in contact with bone spurs when the balloon is inflated in the vertebral body.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER¿ INFLATABLE BONE TAMP
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12785471
MDR Text Key285007520
Report Number9612164-2021-04314
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K041454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/07/2023
Device Model NumberKPT1502
Device Catalogue NumberKPT1502
Device Lot Number222517223
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
-
-