| Catalog Number |
5551000 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, during set-up of a procedure, the bard/davol x-stream controller would not power on.As reported, the device was tested by the biomed who confirmed the device had no power after inserting a tubing cassette used specifically for testing.It is reported that the fuses were good, void seals are intact and there was no damage noted to the device.As reported, there was no patient involvement.Sample evaluation identified heat damage to internal components.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual evaluation of the subject product found thermal damage to internal components (ribbon cable - membrane and u2 logic chip).The heat damage on the chip was located on the surface directly under the ribbon.And while there was space between the chip and the ribbon, heat buildup resulted in thermal damage of the components.The damage to the internal components caused the functional issue reported.The device had been in use for approx.2-years.There was no identification of a manufacturing/material defect.Review of manufacturing records confirms product was manufactured to specification.Sample evaluated.
|
| |
|