STRYKER TRAUMA KIEL TIBIAL NAIL, STANDARD T2 TIBIA Ø10X345 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1822-1034S
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported that during implantation of a t2 tibial nail using the suprapatellar instrumentation the nail appeared to have gotten fused to the nail holding bolt and insertion arm.The staff spent 10-15 minutes attempting to disengage the nail but were unable to.They elected to use the standard approach with a new nail.The nail was able to be disengaged after it was returned to stryker (b)(4).
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.The device inspection revealed the following: the devices were returned in disassembled condition.The received device shows signs of usage evident by scratches on nail body and near the point of contact between the nail holding screw pegs and tibial nail.Overall, the device is in good condition.The received tibia nail was assembled.The nail was in loose condition but engaged in the corresponding pegs of the handle and could be disassembled easily.No indications of material, manufacturing or design related problems were found during the investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿ensure that you are familiar with the intended uses, indications/contraindications, compatibility and correct handling of the implant, which are described in the operative technique manual for the product system.Please remember that product systems may be subject to alterations that affect the compatibility of the implant with other implants or with instruments¿.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by improper handling.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
It was reported that during implantation of a t2 tibial nail using the suprapatellar instrumentation the nail appeared to have gotten fused to the nail holding bolt and insertion arm.The staff spent 10-15 minutes attempting to disengage the nail but were unable to.They elected to use the standard approach with a new nail.The nail was able to be disengaged after it was returned to stryker canada.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL, STANDARD T2 TIBIA Ø10X345 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
sharon rivas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12786411
Report Number0009610622-2021-00782
Device Sequence Number4005139
Product Code HSB
UDI-Device Identifier04546540197917
UDI-Public04546540197917
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K003018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model Number1822-1034S
Device Catalogue Number18221034S
Device Lot NumberK07465B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/13/2021
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date11/10/2021
Supplement Report FDA Received Date02/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1