Model Number 1002717 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the user guide: always make sure that the correct time and date are set on your system.When editing time, always check that the am/pm setting is accurate.No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the customer intermittently experienced low blood glucose (bg 72 mg/dl) and carbohydrates were consumed to address bg.Pump system check was performed with the customer and the pump time was found to be incorrect.Customer reported to have entered the incorrect time during daylight savings.Pump time was corrected.
|
|
Manufacturer Narrative
|
Tandem quality engineer, evaluated pump data.And concluded the following: there is no evidence, that the pump experienced a malfunction or failure.
|
|