Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: visual inspection: the si polyaxl screw 5 x 45mm (p/n: 179712545, lot number: apdbnf) was received at us cq.Visual inspection of the complaint device shows that the shaft of the device was broken and the pc attachment shows the broken embedded device x-ray.Device failure/defect identified? yes.Dimensional inspection: complaint relevant dimensional inspection was not performed due to the geometry of the device.Document/specification review based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Complaint confirmed? yes.Investigation conclusion: this complaint could be confirmed.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: the dhr of product code: 179712545; lot : apdbnf; was electronically reviewed and no non-conformances were observed during the manufacturing process; the product was released on: march 13, 2013.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|