| Model Number |
66000016 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, has been returned for evaluation.There was no obvious visual issue, however, the functional evaluation confirmed the profore layer 4 was difficult to unwind.A relationship against the reported event has been established.A complaint history review confirmed further instances of this nature.A review of the manufacturing records confirmed the device was released according to specification.A further review of the manufacturing process was also performed, and a root cause of inadequate standard operating procedure has been assigned.Corrective action has been assigned regarding this event to reduce the probability of further reoccurrences.This investigation is now complete, smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|
Event or Problem Description
|
|
It was reported that, during set up for treatment, the fourth (4) layer of a profore 18-25cm ankle circ.Case 8 was too fragile, and tore off easily while clinician tried to pull out the layer.Treatment was performed, after a non-significant delay, with a back-up device.Patient was not injured as consequence of this problem.
|
| |
|