MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HFN 10MMX16CM LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HFN 10MMX16CM LEFT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71701016L
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Implant Pain (4561)
Event Date 08/21/2015
Event Type  Injury  
Event Description
It was reported that, after a primary trauma surgery had been performed on the patient's left hindfoot on (b)(6) 2010, in which a hfn 10mm x 16cm left nail was implanted along with three screws placed across the proximal tibial (x1), distal tibial (x1) and distal talar (x1) region to treat a post-traumatic ankle & subtalar arthritis, the patient experienced pain associated to the implanted devices.An additional surgery was performed on (b)(6) 2015 to remove the nail.There is no confirmation on whether this implant was replaced with smith and nephew back up device.The reported incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known and it is not possible to collect it.
 
Manufacturer Narrative
Reference number: case (b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, no relevant supporting clinical information could be provided to assist with this clinical investigation as the data collected from the post-market clinical follow up activity (pmcf) was anonymous.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HFN 10MMX16CM LEFT
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12787456
MDR Text Key280586520
Report Number1020279-2021-07979
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010543127
UDI-Public03596010543127
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71701016L
Device Catalogue Number71701016L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight125 KG
-
-