| Model Number |
PC9000 |
| Medical Device Problem Codes |
No Device Output (1435); Unintended Electrical Shock (4018)
|
| Health Effect - Clinical Codes |
Shock (2072); Electric Shock (2554)
|
| Date of Event |
10/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced electric shock.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced electric shock.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: "it was reported that "(engineer) installed it on the stack first thing this morning, completed an electrical safety test and function check.On pre-use check the theatre staff found that the loan pc9001, sn (b)(6) would not switch on.Initial reporter (staff at account) was called to troubleshoot, as they inspected the device it gave them quite a significant electric shock.They can also confirm it does not switch on."" probable root cause/s: as reported, vpi could not power up.The failure was caused by a power supply failure.The device was returned for investigation and the reported failure mode was reproduced.The reported failure mode will be monitored for future re occurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the user experienced electric shock.
|
| |
|
Additional Manufacturer Narrative
|
|
Updating section d3: manufacturer entity g1: manufacturing site for devices fda registration number to 3012345110 - endoscopy (novadaq).
|
| |
|