| Model Number |
0502704030 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: blurry.Confirmed failure: outer tube damaged (bent, dented).Broken rod lens.Probable root cause: incorrectly assembled optical train, damage to optical train, end of life wear-out, shipping damage, use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|