| Model Number |
3910-400-010 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor broken.Probable root cause: design: anchor not compatible with prepared hole size; anchor thread design makes insertion too difficult; inserter/anchor material/design too weak; anchor tip geometry not sharp enough for application.Process: inserter, implant, manufactured out of specification.Application: excessive force torquing anchor or lateral force applied during insertion; not enough axial force during insertion; use of wrong tap - improperly prepared hole; bone to hard for anchor insertion; bone not hard enough to maintain screw purchase; no pilot hole prepared.The reported failure mode will be monitored for future reoccurrence.H3 other text: 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|