SMITH & NEPHEW, INC. HFN 10MMX16CM RIGHT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
71701016R |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Malunion of Bone (4529)
|
| Date of Event |
01/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reference number: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after a primary trauma surgery had been performed on the patient's right hindfoot on (b)(6) 2015, in which a hfn 10mm x 16cm right nail was implanted along with four screws placed across the proximal tibial (x2), distal transverse (x1) and distal posteroanterior (x1) region to treat a rheumatoid arthritis on the ankle, subtalar and talonavicular joints, the patient experienced an ankle nonunion with a fractured fusion nail implant through the talar screw hole.A revision surgery was performed on (b)(6) 2020 to treat this adverse event.The is no confirmation in regards of the identity of the explanted devices, as well as the devices implanted in exchange.Despite this incident, a subtalar fusion was achieved and confirmed by x-ray assessment on (b)(6) 2019.The reported incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known and it is not possible to collect it.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, no relevant supporting clinical information could be provided to assist with this clinical investigation as the data collected from the post-market clinical follow up activity (pmcf) was anonymous.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Per case details, the reported issue was resolved.No further clinical/medical assessment is warranted at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury,procedural/user error, surgical/post operative complications, healing issues and/or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|