| Model Number |
TRAY DIA. 40 TH 0 ECC 1.5 |
| Medical Device Problem Code |
Inaccurate Information (4051)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device inspection revealed the following: the received device shows that the flex shoulder system reversed tray labelled dia.40 th 0 ecc 1.5 correspond in fact to a flex shoulder system reversed tray th 0 ecc 0.Due to the nature of the complaint, a functional inspection was not considered necessary.Due to the nature of the complaint, a dimensional inspection was not considered necessary.A review of the device history for the reported lot did not indicate any abnormalities.Based on investigation, the root cause was attributed to a manufacturing related issue.A capa was opened and the affected parts were recalled.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
The surgeon asked the nurse to unpack a reversed tray ecc +1.5 (dwf510).However in the packaging was present a reversed tray ecc 0 (dwf500).A new implant have been opened with the same reference but with a different lot.No major prolongation of the surgery.
|
| |
|