Model Number KNEE CAPSULE REPAIR SYSTEM W/GAP GUIDE
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint confirmed.Two unpackaged c15579 gap drill guides from an ar-5875-2 were received for investigation.Visual inspection identified that the c15580 drill guide components from the c15579 gap drill guides had both broken off.One c15580 drill guide fragment was returned, and measurements conducted with digital micrometer id: 224 identified that the component met design specifications.The cause remains undetermined.
 
Event or Problem Description
On (b)(6) 2021, it was reported by a sales representative via email that (2) ar-5875-2 drill guides snapped.This occurred on (b)(6) 2021 during a mensico-capsular repair in conjunction with a meniscal root repair.After the surgeon performed the root portion of the procedure, he had placed the guidewires and began drilling for his suture taks.Upon insertion of the second anchor the gap guide snapped, causing the team to skive with the anchor.A second kit was opened up for the guide, but upon inserting the anchor, the gap guide broke with the anchor staying in the bone.For the third anchor the surgeon chose to freehand the drill instead of opening another kit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE CAPSULE REPAIR SYSTEM W/GAP GUIDE
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12793537
Report Number1220246-2021-03914
Device Sequence Number3947925
Product Code MAI
UDI-Device Identifier00888867307513
UDI-Public00888867307513
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K180594
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberKNEE CAPSULE REPAIR SYSTEM W/GAP GUIDE
Device Catalogue NumberAR-5875-2
Device Lot Number12170195
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2021
Initial Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-