• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS CMV IGG ASSAY; ENZYME LINKED IMMUNOABSORBENT ASSAY, CYTOMEGALOVIRUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS CMV IGG ASSAY; ENZYME LINKED IMMUNOABSORBENT ASSAY, CYTOMEGALOVIRUS Back to Search Results
Catalog Number 07027117190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2021
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter received a questionable elecsys cmv igg result for one patient sample with cobas e 801 module (b)(4).The initial result was 1.20 u/ml (reactive).This result was reported outside of the laboratory and questioned by the doctor.On (b)(6) 2021, the sample was tested in another laboratory on an unknown analyzer with a result of negative (non-reactive).
 
Additional Manufacturer Narrative
The sample was requested for investigation.
 
Additional Manufacturer Narrative
The customer's calibration and qc results were within specification.The customer's sample pre-analytic details were requested but not provided.The result of the customer could be confirmed.Further investigation suggests that the sample is consistent with a false positive in the elecsys cmv igg assay.Based on data presented in product labeling, single false positives can occur.The reagent performs within specification.A general product problem was not found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS CMV IGG ASSAY
Common Device Name
ENZYME LINKED IMMUNOABSORBENT ASSAY, CYTOMEGALOVIRUS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12793606
Report Number1823260-2021-03328
Device Sequence Number5718321
Product Code LFZ
UDI-Device Identifier04015630940172
UDI-Public04015630940172
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K131605
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number07027117190
Device Lot Number568054
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2021
Supplement Date Received by Manufacturer01/21/2022
Initial Report FDA Received Date11/11/2021
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-