• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOVA BIOMEDICAL CORP STATSTRIP GLUCOSE HOSPITAL METER; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOVA BIOMEDICAL CORP STATSTRIP GLUCOSE HOSPITAL METER; BLOOD GLUCOSE METER Back to Search Results
Model Number 54790
Medical Device Problem Code Patient Data Problem (3197)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer is located in (b)(6) where the meter part number is 54794.As the manufacturer, a similar device part number for the us to report is 54790.Therefore, the report is being submitted to the fda under 54790 and the subsequent report to (b)(6) will reflect 54794.The facility reported patient barcode scan error on a patient's identification number.They did not report any patient harm or medical intervention.The facility is continuing to utilize the meter, therefore; no sample was returned to nova for evaluation.A device history record (dhr) review for the meter was performed.The review included an assessment of the production, testing, and release of the products.No abnormalities or concerns were observed; the dhr indicated the released product met all specifications.Previously opened and investigated incidents from the customer facility determined the root cause for the patient misreads is attributed to the user facility using poor quality 1d barcodes that were printed using code -39 symbology.Code-39 is no longer industry standard and has a higher substitution failure rate than code-128 and much higher than the standard 2d barcodes available.This failure mode does not represent a systemic issue or a failure of the meter to perform as intended.The failure mode is directly related to the quality of the barcode being generated by the customer.The root cause of the misread is most likely due to poor barcode quality.Based on the original investigation, nova suggested to prevent barcode mis-reads in the future, the customer should change from a 1d to a 2d barcode, which would greatly reduce the chances of a barcode misread.Trending will be monitored for this and or similar complaints.
 
Event or Problem Description
The customer reported that there was a bar code scanning error on a patient's identification number while using a stat strip (1.86) meter.The incident occurred on (b)(6) 2021.The customer stated that they were unaware of any potential patient harm from the barcode misread and will not be returning the meter as it is currently in use at the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATSTRIP GLUCOSE HOSPITAL METER
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
NOVA BIOMEDICAL CORP
200 prospect street
waltham MA 02454 9141
Manufacturer (Section G)
NOVA BIOMEDICAL
200 prospect street
waltham MA 02454 9141
Manufacturer Contact
shalomy mathew
200 prospect street
waltham, MA 02454-9141
7816473700
MDR Report Key12796602
Report Number1219029-2021-00050
Device Sequence Number12007735
Product Code PZI
UDI-Device Identifier00385480547901
UDI-Public00385480547901
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K181043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number54790
Device Catalogue Number54794
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Initial Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-