Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2021
Event Type  malfunction  
Event Description
It was reported that neonate was cooling on the arctic sun device with 4 hours left till rewarming. The device sounded an alarm, and the therapy was stopped. Esophageal probe was in place. Patient temperature was at the 33. 5c target. Nurse noted that the temperature in cable was frayed and changed out the device. Event log showed alarm 14 (patient temp probe 1 out of range) and alert 114 (treatment stopped). The nurse would order for a new cable. Per follow up information received on 29oct2021, nurse stated that another arctic sun device was used to complete the last two hours of therapy. There were no further issues with the 2nd arctic sun device and therapy was completed successfully.
 
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® TEMPERATURE MANAGEMENT SYSTEM, TEMPERATURE CABLE
Type of DeviceUNKNOWN TEMPERATURE CABLE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12796702
MDR Text Key282977474
Report Number1018233-2021-07214
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 11/11/2021 Patient Sequence Number: 1
-
-