MAUDE Adverse Event Report: TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL LF 6/CS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028484
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
There was a leak in the tubing of the circuit that comes out of the pleurevac which we do not touch (it's wrapped and the only manipulation we do is to connect the end of this tubing with the drain) and which should be totally sealed.Tried to tighten it again but the drained liquid was still leaking out of the tubing.Additional information indicates the leak of pleural fluid occurred at the red/blue connector.
 
Additional Manufacturer Narrative
Qn# (b)(4).The device history record of batch number 74m2000607 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.The device history review shows that the product was assembled and inspected according to our specifications.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE SAHARA DRY SUCT/DRY SEAL LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key12796897
Report Number3004365956-2021-00306
Device Sequence Number20151402
Product Code KDQ
UDI-Device Identifier34026704631668
UDI-Public34026704631668
Combination Product (Y/N)N
Initial Reporter CountryLU
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2023
Device Model NumberIPN028484
Device Catalogue NumberS-1100-08LF
Device Lot Number74M2000607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2021
Initial Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-