MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Gas Output Problem (1266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  malfunction  
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a gas gain in iab circuit alarm.It is unknown under which circumstances the event occurred, however there was no patient involvement, thus no adverse event reported.
 
Manufacturer Narrative
Testing of actual/suspected device(10) a getinge field service engineer (fse) was dispatched to evaluate this unit.Fse found that customer resolved the issue by replacing iab catheter.A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: d5, g1(contact person).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12797264
MDR Text Key280778763
Report Number2249723-2021-02604
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received08/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-