| Model Number |
IPN040969 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "doctor opened the product and put air in it up to 40cc.After then, found leakage on balloon".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that "doctor opened the product and put air in it up to 40cc.After then, found leakage on balloon".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Functional testing was then performed and the cuff and pilot balloon were inflated to 25mbar using a calibrated endotest meter.The cuff could not stay inflated.The sample was then immersed in water for leak testing.Air bubbles were observed coming from the pilot balloon indicating there was a leak.The sample was then observed under a magnifying camera and a tear mark on the side arm pilot balloon was observed.A device history record review was performed and no relevant findings were identified.The complaint of leakage was confirmed; however, a root cause for the issue could not be determined.There is 100% inflation and leak testing conducted during the manufacturing process whereby any leakage would be detected prior to release from the manufacturing facility.
|
| |
|