• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS Back to Search Results
Model Number 521-07-238
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as glenoid loosening.The previous surgery and the surgery detailed in this event occurred 1 year and 1 month apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to glenoid loosening.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to glenoid loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE ANATOMIC SHOULDER
Common Device Name
ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 38, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key12798737
Report Number1644408-2021-01237
Device Sequence Number14641863
Product Code KWS
UDI-Device Identifier00190446178354
UDI-Public(01)00190446178354
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/09/2024
Device Model Number521-07-238
Device Catalogue Number521-07-238
Device Lot Number890U1059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2021
Initial Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-07-000 LOT 888U1460; 520-46-316 LOT 957U1082
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
-
-