• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVE DIVISION HANCOCK II MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVE DIVISION HANCOCK II MITRAL BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number T510C27
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Tricuspid Valve Stenosis (2113); Tricuspid Valve Insufficiency/ Regurgitation (4453); Insufficient Information (4580)
Date of Event 11/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: the device remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 3 years and 4 months post implant of this 27mm aortic bioprosthetic valve implanted in the pulmonary position, a transcatheter valve was implanted valve-in-valve.The reason for intervention was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that corrects the previous valve location for the valve replacement.Three years and four months prior to this surgery, the patient had a 27 mm bioprosthetic mitral valve implanted in the tricuspid and an aortic bioprosthetic valve in the pulmonary position.The surgeon reported that the the pulmonary position was functioning well.No interventions were needed on the valve in the pulmonary position.The valve in the tricuspid position has developed mild to moderate stenosis and severe regurgitation.The transcatheter pulmonary valve was implanted inside the bioprosthesis in the tricuspid position.The surgeon reported, "the patient did well after the procedure but did have an anaphylactic reaction to erythromycin eye ointment given the evening of the procedure for mild corneal abrasion.She recovered well from this.".No additional adverse patient effects were reported.A4: weight updated h6: codes updated d1: updated d4: updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANCOCK II MITRAL BIOPROSTHETIC HEART VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key12799169
Report Number2025587-2021-03415
Device Sequence Number5111860
Product Code LWR
UDI-Device Identifier00643169594326
UDI-Public00643169594326
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P980043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/27/2020
Device Model NumberT510C27
Device Catalogue NumberT510C27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2021
Supplement Date Received by Manufacturer11/29/2021
Initial Report FDA Received Date11/11/2021
Supplement Report FDA Received Date12/02/2021
Date Device Manufactured08/28/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age37 YR
Patient SexFemale
Patient Weight58 KG
-
-