| Model Number |
15-176 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: crease fold, crease fold, white particles on the surface of the shell, weight to the spec.Cloudy was observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.The device has been explanted.
|
| |
|