• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE-RITE PREVUE PLUS; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE-RITE PREVUE PLUS; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770086
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of the information, the following was concluded: observation found during evaluation: there is a dark vertical line in the ultrasound image due to an internal failure within the probe.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Observation found during evaluation: there is a dark vertical line in the ultrasound image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE PREVUE PLUS
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key12800638
Report Number3006260740-2021-04827
Device Sequence Number4543116
Product Code IYO
UDI-Device Identifier00801741095450
UDI-Public(01)00801741095450
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K120882
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770086
Device Catalogue Number9770086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2021
Initial Report FDA Received Date11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-